Pilot validation study of the PIMA portatil platform for the CD4 T-cell enumeration compared to flux cytometry

Authors

  • Omar Sued
  • Pablo Salgado
  • Camilia Picchio
  • María I. Figueroa
  • María E. Socías
  • Osvaldo Cando
  • Liliana Díaz
  • Ricardo Hermes
  • Héctor Peréz
  • Pedro Cahn

DOI:

https://doi.org/10.52226/revista.v23i87.256

Keywords:

PIMA, Flow Citometry, CD4 count

Abstract

Objective: To compare PIMA methodology to the standard CD4 FACSCalibur flow cytometry for CD4 testing.
Method: Paired blood samples were collectedamong HIV patients and tested using PIMA and Becton Dickinson FACSCalibur.
Results: 223 samples were studied in parallel. There was a high concordance, being Pearson Linear Correlation 0.974, the Spearman Rank Correlation Coefficient 0.971. The Intraclass Correlation Coefficient for individual measures was 0.965 (95%CI 0.926-0.980) and for average measures 0.982 (95%CI 0.961-0.990). There was a low coefficient of variation from duplicate measurements (11.4%).
Conclusion: This study shows a good correlation between the PIMA CD4 count and the Becton Dickinson FACSCalibur. By being a point of care methodology that produces same-day results, PIMA CD4 might be an alternative for sites without access to standard CD4 count methodology.

Downloads

Download data is not yet available.

Author Biographies

  • Omar Sued, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

  • Pablo Salgado, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

  • Camilia Picchio, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

  • María I. Figueroa, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

  • María E. Socías, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

  • Osvaldo Cando, ,

    Servicio de Laboratorio. H.G.A. Juan A. Fernández. Buenos Aires, Argentina.

  • Liliana Díaz, ,

    Servicio de Laboratorio. H.G.A. Juan A. Fernández. Buenos Aires, Argentina.

  • Ricardo Hermes, ,

    Servicio de Laboratorio. H.G.A. Juan A. Fernández. Buenos Aires, Argentina.

  • Héctor Peréz, ,

    División de Infectología. H.G.A. Juan A. Fernández. Buenos Aires, Argentina.

  • Pedro Cahn, ,

    Área de Investigaciones Clínicas. Fundación Huésped. Buenos Aires, Argentina.

    División de Infectología. H.G.A. Juan A. Fernández. Buenos Aires, Argentina.

Downloads

Published

2023-10-09

Issue

Section

Original article

How to Cite

Pilot validation study of the PIMA portatil platform for the CD4 T-cell enumeration compared to flux cytometry. (2023). Actualizaciones En Sida E Infectología, 23(87). https://doi.org/10.52226/revista.v23i87.256